Warning: Exceeds Max Basal
Cause
Basal delivery rate (or Temp Basal delivery) exceeds
user-set maximum.
Effect Basal delivery stops.
Message
Displayed when awakened (by basal delivery attempt
every 3 min. or manually) until confirmed.
Action
Press OK to confirm. Reprogram Max Basal amount in
Setup Advanced (or reprogram Temp Basal).
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Delivery Canceled due to Low Cartridge
Cause
Basal or Bolus delivery exceeds insulin remaining in
cartridge.
Effect Basal or Bolus delivery stopped.
Message
Once per occurrence and each time awakened until
confirmed.
Action Press OK to confirm. Replace with full cartridge.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: No Prime, No Delivery
Cause Pump is not primed.
Effect All deliveries stop.
Message Every 3 minutes or when awakened manually.
Action Press OK to confirm. Disconnect, reprime.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Alarm: Occlusion
Cause Occlusion detected.
Effect All deliveries stop.
Message Continuous until confirmed.
Action
Press OK to confirm. Disconnect and prime to clear
occlusion. Option to select Suspend (see Suspend).
Beeps/Vib
User-selected, every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour. (Once
confirmed, No Prime warning triggered, see No Prime.)
Alerts, Warnings and Alarms