1. From Basal Menu, scroll to Temp. Press OK.
2. The Change % field will flash to indicate Edit mode. Use the Up/Down arrows to enter the
percentage change desired. Press OK button to exit Edit mode.
3. Duration field is highlighted. Press OK to activate Edit mode.
4. Use the Up/Down arrows to enter the duration desired. Press OK to exit Edit mode.
5. Go is highlighted. Press OK to activate Temp Basal.
The lowest basal delivery amount possible is 0.025 U/hr. When you set a negative
temporary basal rate, the pump will beep and display an Alert screen to remind you
of the minimum delivery limit. This screen will display once for 5 seconds and give
one audible alert (if you turned on Alert sounds in Setup).